Vaisala HUMICAP HMT330 SERIES Uživatelská příručka Strana 133

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 178
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 132
Chapter 5 ______________________________________________________________ Maintenance
VAISALA ______________________________________________________________________ 133
Error States
In error state the quantity is not measured and the output is shown as
follows:
- Analog channel outputs 0 mA or 0 V (you can use the serial line
command AERR or display/keypad to change this fault indication
value, see section Analog Output Fault Indication Setting on page
113.)
- The serial port outputs stars (***)
- The cover LED is blinking
- Optional display: error indicator is lit.
0706-008
Figure 69 Error Indicator and Error Message
The following number refers to Figure 69 above:
1 = Error Indicator
The error indicator disappears when the error state is over and you have
checked the error message. Press the INFO button to display the error
message.
You can also check the error message via the serial interface by using the
command ERRS. In case of constant error, please contact Vaisala, see
Vaisala Service Centers on page 136.